Experimental evaluation of central venous pressure monitoring in the dog
The measurement of central venous pressure (CVP) will provide objective information about intravascular blood volume. Limitations exist with direct and indirect methods of measuring CVP; however, the purpose of this study was to determine if such factors as catheter size, patient position, or measurements taken with either a water manometer or a strain-gauge transducer would affect the accuracy of CVP measurements. The study population consisted of six adult, mixed-breed dogs. Catheter sizes evaluated were 16- and 19-gauge. Patient positions evaluated were right and left lateral recumbencies. Right atrial pressures (RAPs) were used as the reference standards for CVP measurements. Measurements of RAP and CVP were obtained with a water manometer and a strain-gauge transducer. No significant differences were observed in measured CVPs for the effects of catheter size and right versus left lateral recumbency. A 2.7-mmHg difference in water manometry was demonstrated when compared to strain-gauge transducer measurements. Also, strain-gauge transducer measurements had less precision compared to water manometer measurements.